- Kit/Kits
- 1
- Kit/Kits
- 1283 INR
- Industrial
- Industrial Lab Chemicals
- 95
- Auto zyme Bilirubin is a reagent set for determination of Total and Direct Bilirubin based on Jurassic & Grof method, using Diazotized Sulphonic Acid with DMSO as an activator for Total Bilirubin.
- Auto Zyme Bilirubin is a two reagent system using one step procedure.
- Auto Zyme Bilirubin is linear upto 20 mg%.
- Auto Zyme Bilirubin can be used on any Colorimeter, Spectrophotometer, Discrete semi-automated and Automated analyzer. Programmed can be designed for any specific analyzer upon request.
- Auto Zyme Bilirubin has one step reconstitution.
- Auto Zyme Bilirubin can be determined in just 5 minutes.
- Auto Zyme Bilirubin method is relatively free of interference from hemoglobin and other commonly occurring substances in serum or plasma.
- This insert covers technical matter for the complete range of Auto Zyme Bilirubin kits available viz.:
Product | Pack-Size |
Bilirubin T & D (Total and Direct) 200 | 1 x 200 ml |
Bilirubin Total 1000 | 1 x 1000 ml |
Bilirubin Direct 1000 | 1 x 1000 ml |
Please read the appropriate matter depending on the kit you have purchased.
PRINCIPLE
Bilirubin reacts with diazotized sulphanilic acid to produce azobilirubin (violet colour). DMSO catalyzes the formation of azobilirubin from free bilirubin. The violet colour is proportional to bilirubin concentration measured at 546 nm.(530 - 550 nm.).
TOTAL BILIRUBIN
Bilirubin + Sulphanilic Acid + Sod. Nitrite DMSO x p Azobilirubin
DIRECT BILIRUBIN
Bilirubin + Sulphanilic Acid + Sod. Nitrite .Azobilirubin
DMSO Dimethyl Sulphoxide
REAGENT STORAGE & STABILITY
- The reagent kit should be stored at Room Temperature (25-30 degree C) and is stable till the expiry date indicated on the label.
- The working solution is stable for 2 days at 2-8 degree C and for 5 hours at Room Temperature (25-30 degree C).
Component | Concentration | |
| Total Bilirubin | Direct Bilirubin |
Sulphanilic Acid | 32 mmo1/1 | 32 mmo1/1 |
Sodium Nitrite | 290 mmo1/1 | 29 mmo1/1 |
Hydrochloric Acid | 165 mmo1/1 | 165 mmo1/1 |
DMSO | 7 mmo1/1 |
|
SPECIMEN COLLECTION & PRESERVATION
Blood should be collected in a clean dry container. Although serum is preferred plasma can also be used as sample.
Following anticoagulant can be used for plasma separation :
- EDTA 2 mg/ml of blood
- HEPARIN 200 IU/ml of blood
Bilirubin is light sensitive. Avoid exposure of serum or plasma to direct light. Bilirubin in serum and plasma is stable for a day at 2-8 degree C or one month at -10 degree C. Samples should be brought to room temperature before use.
PROCEDURE
- Reaction type: End-Point
- Reaction time: 5 mins. at R.T. (25-30 degree C)
- Wavelength: 546 nm.(530 - 550 nm.)
- Zero setting with : Serum Blank
- Sample volume: 0.05 ml (50 III)
- Reagent volume: 1.02 ml
- Factor: 20.2
- Linearity: 20 mg%
Perform the assay as given below :
Serum Blank | | Test |
Reagent T1 | 1.0 ml | 1.0 ml |
Reagent T2 |
| 0.02 ml |
Serum / Plasma | 0.05 ml | 0.05 ml |
DIRECT BILIRUBIN
Serum Blank |
| Test |
Reagent T1 | 1.0 ml | 1.0 ml |
Reagent T2 |
| 0.02 ml |
Serum / Plasma | 0.05 ml | 0.05 ml |
Mix, and incubate the assay mixture at Room Temperature (25-30 degree C) for 5 minutes. Read before 8 minutes the absorbance of the test against their respective blanks at 546 nm (530 - 550 nm).
EXPECTED VALUES
- Total Bilirubin: up to 1.0 mg%
- Direct Bilirubin: up to 0.3 mg%
- EXW ABPL
- Cash Against Delivery (CAD), Cash in Advance (CID), Cheque, Letter of Credit (L/C), Letter of Credit at Sight (Sight L/C)
- 10000 Per Day
- 2 Days
- Yes
- packaging : Empity
- Asia, Australia, Central America, North America, South America, Eastern Europe, Western Europe, Middle East, Africa
- All India
- certification : iso
212, Udyog Mandir No. 17/C, Bhagoji Keer Marg, Mahim, Mumbai, Maharashtra, 400016, India
Phone :08045479256