- Infinite HDL-C is a reagent set for direct quantitative determination of HDL-Cholesterol in human serum or plasma based on Direct Homogenous method.
- Direct measurement by .finite HDL-C overcomes the limitation of precipitation method thereby allowing accurate determination of HDL-C.
- Infinite HDL-C is a ready-to-use, two liquid reagent system stable till expiry.
- Infinite HDL-C is linear upto 185 mg/dl.
- There is no significant interference in hyper-lipemic and hyper-immunoglobulinemic samples.
- Results obtained with Infinite HDL-C show excellent correlation with ultracentrifugation method.
Lyophilized calibrator should be reconstituted by adding 1.0 ml of distilled or deionized water. Dissolve the contents of the vial by swirling gently to avoid the formation of foam. Do not shake. Let stand for 20 minutes at room temperature before use.
REAGENT STORAGE STABILITY & HANDLING
- The reagent kit should be stored at 2 - 8 degree C and is stable till the expiry date indicated on the label.
- Reconstituted calibrator is stable for 3 days at 2 degree - 8 degree C. Reconstitution stability of the calibrator may be extended up to 30 days by aliquoting and freezing the reconstituted calibrator preparation at less than -20 degree C.
- Reaction type : End Point (2 Step)
- Reaction time : 5 + 5 mins.
- Wavelength: 578 nm.
- Flowcell temperature: 37 degree C
- Zero setting with: Distilled water
- Sample volume: 0.01 ml (10 NI)
- Reagent volume: 0.750 ml + 0.250 ml
- Linearity : 185 mg/dl
Perform the assay as given below:
| Calibrator | Sample |
R1 | 0.01 ml | 0.01 ml |
| 0.750 ml | 0.750 ml |
- Mix and incubate for 5 minutes at 37 degree C.
- Measure the absorbance Al at 578 nm.
| Calibrator | Sample |
R2 | 0.250 ml | 0.250 ml |
- Men : 30 - 70 mg/di
- Women : 30 - 85 mg/di
Expected range varies from population to population. It is therefore recommended that each laboratory should establish its own normal range.
SPECIMEN COLLECTION & PRESERVATION
Blood should be collected in a clean and dry container. Although serum is preferred, anticoagulants such as heparin, citrate, oxalate, EDTA and sodium fluoride do not have significant influence on the assay when they are used in their usual amounts. Specimen analysis should be done immediately after collection. Do not use the specimens that are repeatedly subjected to freezing and thawing.
PROCEDURE LIMITATIONS
If the HDL-C value exceeds the linearity limit of 185 mg/dl, then dilute the specimen suitably with normal saline and repeat the assay. In such case the assay value should be multiplied with the dilution factor to obtain the correct HDL-C value of the specimen.
QUALITY CONTROL
To ensure adequate quality control, it is recommended that each batch should include a normal and an abnormal commercial reference control serum. It should be realized that the use of quality control material checks both instrument and reagent functions together. Factors which might affect the performance of this test include instrument function, temperature control, cleanliness of glassware and accuracy of pipetting.
- EXB ABPL
- Cash Against Delivery (CAD), Cash on Delivery (COD), Cash Advance (CA), Cash in Advance (CID), Cheque, Delivery Point (DP), Paypal
- 10000000 Per Day
- 2 Days
- Free samples available with shipping and taxes paid by the buyer
- Packaging: Empity
- Asia, Australia, Central America, North America, South America, Eastern Europe, Western Europe, Middle East, Africa
- All India
- ISO
212, Udyog Mandir No. 17/C, Bhagoji Keer Marg, Mahim, Mumbai, Maharashtra, 400016, India
Phone :+918045479256